Web-IO CO2
Monitor air quality
Web-IO CO2 is a network-capable measuring instrument for detecting carbon dioxide concentration in the ambient air.
For long-term, continuous detection of the CO2 concentrations it is equipped with an internal data logger, and the measuring data can be sent to the W&T cloud as well. The alarm function sends messages when configurable limits are exceeded.
The measuring range of 0-5,000 ppm is suitable for monitoring the indoor air in schools, daycare centers, retirement homes and office buildings. Applications from agriculture and horticulture include stall climate detection and greenhouse monitoring.
Properties
Sensor:
- CO2 sensor:
- Measuring range: 0-5000 ppm CO2
Connectivity:
- Configurable LED tower on sensor for optical indication of limit values
- Use browser to monitor CO2 and trends
- Alarm and reporting function:
- E-mail for alarm or reporting functions
- SNMP polling / alarm traps
- Configure up to 12 alarm messages
- Dynamic integration into other Web sites:
- Direct access to current measurement values, e.g. JavaScript (AJAX).
- Current Industry 4.0 protocols:
- OPC UA, REST and MQTT support
- Additional software interfaces for incorporating into your systems/databases:
- SNMPv1, SNMPv2c, SNMPv3
- Modbus-TCP
- OPC UA, OPC DA
- Syslog
- Sensobase (database integration via ODBC)
- TCP and UDP sockets, client and server
- FTP (data logging)
- Possible applications:
- Monitor CO2 values in the office
- Logging of measurements via FTP, Excel file, email attachment and internal memory
Data logger:
- Internal data logger
- Memory capacity: min. 7 weeks, max. 20 years
- Save frequency: 15, 30 sec, 1, 5, 15, 60 min
- Document measurement data online in the W&T Cloud and access from anywhere in the world
- Internal clock
- Time synchronization using time server calibration
- Battery-backed device clock
Standards & more
- Supply voltage via Power-over-Ethernet (PoE)
- Phantom power using data pairs
- Power over unused wire pairs
- External power supply is an alternative
- Conforms to standards both in office and industrial environments:
- High noise resistance per EN 61000-6-2
- Low noise emission per EN 55032:2015 + A1 Cl. B, EN 61000-3-2 & EN 61000-3-3
- 5 year guarantee
Technical data
Connections and displays:
- Sensor:
- CO2 air quality measurement
- DB9 receptacle
- Cable length: 2m (can be extended to max. 20m)
- Additional temperature and relative humidity measurement on request
- Network:
- 10/100BaseT Autosensing/Auto-MDIX
- RJ45
- IPv6 on request
- Galvanic isolation:
- Network connection min. 1500 V
- Supply voltage:
- Power-over-Ethernet (PoE) or
- DC 12V .. 48V (+/-10%)
- Supply connection:
- Plug-in screw terminal, 5.08mm spacing
- Labeled "L+" and "M"
- Current consumption:
- PoE Class 1 (0.44 - 3.84W)
- typ. 100mA @ 24V DC with external supply
- Displays:
- 1 LED power
- 2 LEDs network status
- 4 LEDs status and error
- 3 LEDs traffic-light display in sensor
Measuring unit:
- Measuring range:
- 0-5000 ppm CO2
- Resolution:
- 1 ppm CO2
- Measuring error:
- Air quality (0-5000 ppm):
- max. ±50 ppm +3% of the measured value
Measuring frequency:
4s
- Storage frequency:
- 15s, 30s, 1m, 5m, 15m, 60m
- Memory depth (4MB):
- min. 7 weeks, max. 20 years
Housing and other data:
- Housing:
- Plastic compact housing for top-hat rail mount
- 105x22x75mm (LxWxH)
- Enclosure rating:
- IP20
- Weight:
- approx. 200g
- Ambient temperature
- Storage: -40..+70°C
- Operating: 0 .. +50°C
- Permissible relative humidity:
- 0...90% relative humidity (non-condensing)
- Scope of delivery:
- 1x Web-IO CO2 for DIN rail mount
- 1x W&T sensor, 2m (CO2)
- 1x Quick Guide
* Our offering is intended only for commercial users. We will be happy to refer private end customers to trading partners through whom our products can be purchased.